Image of PENGARUH DIAMETER PORI TERHADAP PERMEABILITAS ALIRAN BIOFLUIDA PADA PERMUKAAN MEMBRAN PERMEABEL

Tugas Akhir Mesin

PENGARUH DIAMETER PORI TERHADAP PERMEABILITAS ALIRAN BIOFLUIDA PADA PERMUKAAN MEMBRAN PERMEABEL

Muhammad Agra Wicaksono - Nama Orang;

he performance of a dialyzer in a hemodialysis device is strongly influenced by permeable membrane pores. Research on the influence of pores on mass transfer is relatively limited. In this case the study was conducted to vary the influence of pore diameter with sizes of 0.4 μm and 3 μm. Tests were carried out to determine the mass transfer, shear stress, Reynold number, pressure drop, and trans membrane pressure using NaCl and NaOH biofluids. The test is carried out for 5 minutes with a flow rate of 20 ml / minute. The results showed that the greater the pore diameter of the membrane will cause the greater mass transfer that occurs. In testing with NaCl with a pore diameter of 0.4 μm and 3 μm the diffusion coefficients were obtained respectively 4,6x10-8 m3/s and 8,43x10-8 m3/s. The same thing was obtained in testing with NaOH at diameter of 0.4 μm and 3 μm the diffusion coefficients were obtained respectively 5,61x10-8 m3/s and 6,87x10-8 m3/s. In NaOH test was obtained that the shear stess, pressure drop, and Reynold numberthat the bigger the pore diameter, then the number obtained from the test will be smaller. Inversely proportional to testing the smaller the pore diameter, then the calculation result is bigger. In the NaCl testing was also obtained the same result as NaOH, but the NaCl score is higher than NaOH


Ketersediaan
2019219616.151;Muh;Illus; 21x29 cmTersedia
Informasi Detail
Judul Seri
-
No. Panggil
616.151;Muh;Illus; 21x29 cm
Penerbit
Universitas Mataram : Fakultas Teknik Unram., 2019
Deskripsi Fisik
xvii; 67 Hlm; Illus; 21x29 cm
Bahasa
Indonesia
ISBN/ISSN
-
Klasifikasi
616.151
Tipe Isi
other
Tipe Media
computer
Tipe Pembawa
audio disc
Edisi
-
Subjek
Eritrosit, kelainan Talasemia
dialyzer mass transfer pore membrane shear stre
pressure drop Reynold number
Info Detail Spesifik
PENGARUH DIAMETER PORI TERHADAP PERMEABILITAS ALIRAN BIOFLUIDA PADA PERMUKAAN
Pernyataan Tanggungjawab
Muhammad Agra Wicaksono
Versi lain/terkait

Tidak tersedia versi lain

Lampiran Berkas
  • PENGARUH DIAMETER PORI TERHADAP PERMEABILITAS ALIRAN BIOFLUIDA PADA PERMUKAAN MEMBRAN PERMEABEL
Komentar

Anda harus login sebelum memberikan komentar

Perpustakaan Fakultas Teknik
  • Informasi
  • Layanan
  • Pustakawan
  • Area Anggota

Tentang Kami

Ruang Baca Fakultas Teknik

Kami menyedian Buku ajar ilmu teknik dan Tugas Akhir mahaiswa Fakultas Teknik Universitas Mataram

Cari

masukkan satu atau lebih kata kunci dari judul, pengarang, atau subjek

Donasi untuk SLiMS Kontribusi untuk SLiMS?

© 2025 — Senayan Developer Community

Ditenagai oleh SLiMS
Pilih subjek yang menarik bagi Anda
  • Karya Umum
  • Filsafat
  • Agama
  • Ilmu-ilmu Sosial
  • Bahasa
  • Ilmu-ilmu Murni
  • Ilmu-ilmu Terapan
  • Kesenian, Hiburan, dan Olahraga
  • Kesusastraan
  • Geografi dan Sejarah
Icons made by Freepik from www.flaticon.com
Pencarian Spesifik